Mostrar el registro sencillo del ítem

dc.contributor.authorMENDOZA, ZENON ANDRES CUCHO
dc.contributor.authorGALLEGOS, EDUARDO NELSON CHAVEZ
dc.contributor.authorAGUILAR, JOSE ANTONIO FARFAN
dc.contributor.authorROJAS, YESMI KATIA ORTEGA
dc.contributor.authorGOMERO, MILTHER MAX SIFUENTES
dc.contributor.authorOSTOS, NESTOR GOMERO
dc.date.accessioned2025-02-28T20:39:55Z
dc.date.available2025-02-28T20:39:55Z
dc.date.issued2024
dc.identifier.urihttps://hdl.handle.net/20.500.12952/9905
dc.description.abstractTHE WORLD HEALTH ORGANIZATION MENTIONS THAT THE MORTALITY OF PREMATURE NEWBORNS VARIES WIDELY BETWEEN 10% AND 90% DEPENDING ON THE COUNTRY AND ECONOMIC INCOME, OBSERVING THAT HIGH-INCOME COUNTRIES HAVE LOWER MORTALITY RATES IN NEWBORNS BORN PREMATURELY, DUE TO ACCESS TO ADVANCED MEDICAL RESOURCES AND BETTER NEONATAL MEDICAL CARE INFRASTRUCTURE. THEREFORE, IT IS IMPERATIVE TO RESEARCH AND USE CURRENT TECHNOLOGY FOR THE MANUFACTURE OF AFFORDABLE INCUBATORS IN OUR COUNTRY TO SAVE THE LIVES OF PREMATURE NEWBORN BABIES. A NEW NEONATAL INCUBATOR SYSTEM WITH THREE LAYERS OF INSULATION HAS BEEN DEVELOPED AND PATENTED IN THIS RESEARCH. BASED ON A 32-BIT MICROCONTROLLER AT91SAM3X8E OF THE ARDUINO DUE MODULE AS THE ELECTRONIC SYSTEM OF THE PROTOTYPE AND THE RTD-PT1000 DEVICE WAS USED TO READ THE SYSTEM. ACCORDING TO THE PROTOCOL OF THE IEC 60601-2-19 STANDARD, EXPERIMENTAL TESTS FOR NEONATAL INCUBATORS WERE CARRIED OUT. THE METHODOLOGY USED FOR THE ANALYSIS AND EVALUATION MADE IT POSSIBLE TO DEMONSTRATE THE THERMAL EFFICIENCY IN THE PROTOTYPE OF THREE INSULATING LAYERS BY OBTAINING RESULTS WITH AN ERROR LOWER THAN THOSE ESTABLISHED BY THE STANDARD. © 2024 IEEE.
dc.formatapplication/pdf
dc.language.isospa
dc.publisher2024 7TH INTERNATIONAL CONFERENCE ON ELECTRICAL ENGINEERING AND GREEN ENERGY, CEEGE 2024
dc.rightsinfo:eu-repo/semantics/openAccess
dc.rights.urihttps://creativecommons.org/licenses/by/4.0/es_PE
dc.subjectINCUBATOR, PREMATURE NEWBORN, THERMAL EFFICIENCYes_PE
dc.titleTEMPERATURE CONTROL IN A 3-LAYER INSULATING NEONATAL ENCLOSURE TO IMPROVE THERMAL EFFICIENCYes_PE
dc.typeinfo:eu-repo/semantics/articlees_PE
dc.identifier.doi10.1109/CEEGE62093.2024.10744081
dc.subject.ocdehttps://purl.org/pe-repo/ocde/ford#2.00.00es_PE